AfterMath Quick Start: Cyclic Voltammetry
| Step 1: Make sure your potentiostat is connected to the computer system and is turned on. |
|---|
| Step 2: Login to the AfterMath software (which you should have previously installed on your computer). Installation instructions may be found at the link below: http://www.voltammetry.net/pine/aftermath/install |
|
| Step 3: Physically connect the potentiostat to your electrochemical cell using an appropriate cell cable. Note that some Pine potentiostats come with a “Dummy Cell” as shown below that may be used for quick verification of the instrument and software. |
| Step 4: Examine the Instrument Status (see below). Initially, the status should indicate that the cell is a “disconnected” state. If desired, you may use the controls to apply a known idle condition to the cell. In the example below, the instrument has been adjusted to idle in the potentiostat mode while applying +1.2 volts to the working electrode. |
|
| Step 5: From the AfterMath home page, select the “Cyclic Voltammetry” option from the experiment list (see below). (Alternately, you may choose “Cyclic Voltammetry” from the “Experiment” menu.) A new cyclic voltammetry experiment specification is created and placed in a new archive. |
|
| Step 6: Enter the required parameters describing the cyclic voltammetry experiment into the boxes which are shaded yellow. To use a set of default parameters, click on the “I Feel Lucky” button. |
|
| Step 7: Choose your potentiostat in the drop-down menu (to the left of the “Audit” button). Then, press the “Perform” button to start the experiment. |
|
| Step 8: Monitor the progress of the experiment on the real time plot or the progress bar. |
|
| Step 9: The results of the experiment are placed in a study folder in the archive. In addition to the main plot of the voltammogram, additional graphs are created in the “Other Plots” folder. The results are also available in tabular form. |
|
| NOTE: The diagonal line in the plot above is clearly not an actual voltammogram. This pure ohmic response reflects the summed value of two series resistors on “Dummy Cell C” as follows: The charge transfer resistor (Rct) is nominally 1000 ohms, and the uncompensated resistor (Ru) is 100 ohms. Thus, the slope of the diagnonal line reflects a combined nominal resistance of 1100 ohms. (The precise value of the summed resistance can be deduced from the inverse slope of the diagonal line. In this case the Rct + Ru value is 1099 ohms.) |